Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus
Mounting
Hauser
13
Probe Ø 16 mm (0.63 in)
• a < Ø 26 mm (1.02 in)
• b = Ø 16.5 mm (0.65 in)
NOTICE
Poor grounding of the probe end may cause incorrect measurements.
‣
Use a sleeve with a narrow bore to ensure good electrical contact between the sleeve and
the probe rod.
NOTICE
Welding can damage the main electronics module.
‣
Before welding: Ground the probe rod and remove the electronics.
Securing coax probes
For WHG approval: a support is required for probe lengths ≥ 3 m (10 ft).
A0012608
Coax probes can be secured (fixed) at any point in the ground tube.
5.1.3
Shortening the probe
See Operating Instructions.