Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus
Mounting
Hauser
15
*
A0012852
Pay attention to the following when introducing the rope probe into the vessel:
• Uncoil the rope slowly and lower it carefully into the vessel.
• Make sure the rope does not bend or buckle.
• Avoid uncontrolled swinging of the weight, as this could damage internal fittings in the
vessel.
5.2.4
Turning the transmitter housing
To provide easier access to the connection compartment or display module, the transmitter
housing can be turned:
max. 350°
8 mm
8 mm
2.
3.
1.
A0032242
1. Unscrew the securing screw using an open-ended wrench.
2. Rotate the housing in the desired direction.
3. Tighten the securing screw (1.5 Nm for plastic housing; 2.5 Nm for aluminum or
stainless steel housing).