Technical data
Flowfit CYA27
46
Hauser
12.4 Process
Process temperature range
0 to 60 °C (32 to 140 °F), non-freezing
Process pressure range
0 to 4 bar (0 to 58 psi) relative
Pressure/temperature
ratings
0
15
29
44
58
32
50
68
86
104
122
140
158
0
1
2
3
4
0
10
20
30
40
50
60
70 T[°C]
T[°F]
p [bar]
p [psi]
A
B
A0044367
10
Pressure/temperature ratings
A
Process pressure
B
Medium temperature
pH range
pH1 to 12
Process connections
G 1/4" (ISO 228)
Flow
Recommended flow range
5 l version
5 to 8 l/h (1.32 to 2.11 gal/h)
30 l version
30 to 40 l/h (7.92 to 10.46 gal/h)
Critical upper limits
5 l version
40 l/h (10.56 gal/h)
30 l version
80 l/h (21.13 gal/h)
Above the specified flow rate, the pressure in the assembly may exceed the
specification limits of the sensors.