Flowfit CYA27
Installation
Hauser
27
1. The assembly is supplied to the customer with a dummy screw mounted in the
assembly.
2. Remove dummy screw, thrust collar and O-ring (1) from the assembly using a
hexagon wrench AF17.
3. Slide the sensor into the adapter of the assembly.
4. Screw the sensor into the adapter of the assembly.
The CLS82D must not be installed in the pH or oxygen module, as measured errors
may occur due to the short distance to the wall.
5.6
Connecting optional accessories
Option
Process adapter
QA
G 1/2 (ISO 228-1)
A0043724
QB
G 1/8 (ISO 228-1)
A0043723
QH
NPT 1/4"
A0043722
QG
NPT 1/2"
A0043721
QM
Hose connector OD 6 mm (0.24 in), ID 4 mm (0.16 in)
A0043720
QN
Hose connector OD 8 mm (0.31 in), ID 6 mm (0.24 in)
A0043719
PC
Adapter PAL G 1/4
A0043718
The process adapters are supplied with an O-ring seal on the assembly side.
5.7
Post-installation check
1. Close any sampling valves that may be installed.
2. Open needle valves to control flow at the assembly.
3. Close any pressure-reducing valves that may installed upstream from the assembly.
4. After installation, check all connections to ensure they are correctly installed, secure
and leak-tight.
5. Check all pipes and hoses for damage.