CUM 750 / CUS 70
Operation
Hauser
31
Setpoint:
Set the switch point of the relay.
Setting range: 0 ... 100m
Dead band:
Set the value at which the relay status should be maintained until the measured variable has changed
by the hysteresis value.
Setting range: 0 ... 100m
Timer on:
This function provides each relay with a timer which controls sensor rinsing.
Cycle:
Set the rinse interval.
Setting range: 5 ... 2880min (max. 2 days)
Duration:
Set the rinse period.
Setting range: 2 ... 1800s (max. 30 mins.)
Operation:
Move through the submenu by pressing the
→
→
→
→
key, select numeric values by pressing the '
↑↑↑↑
‘ / '
↓↓↓↓
‘ keys
(see Chap. 6.1 "Key assignment"). Confirm and return to the main menu by pressing the ‘E’ key.
6.7.3 SERIAL PORTS submenu
Define the communication parameters for the serial interface using this menu.
Baud rate:
Set the transfer rate.
Setting range: 1200 ... 56,000 bytes
Output:
Select the configuration type for the serial interface of the communication device.
Setting range: ASCII, BINARY
RS 232 / 485:
Select the type of serial communication interface.
Setting range: RS 232 / RS 485
Operation:
Move through the submenu by pressing the
→
→
→
→
key, select numeric values by pressing the '
↑↑↑↑
‘ / '
↓↓↓↓
‘ keys
(see Chap. 6.1 "Key assignment"). Confirm and return to the main menu by pressing the ‘E’ key.