INTENSOR electronics
(operation using
Commulog VU 260 Z)
Note!
The "Zero Correction", "Editing Limits" and "Alarm Current MAX" functions are not included
in the INTENSOR electronice (software version 5.0).
SW/BA
Device
and
Software
No.
VU 260 Z
Changes
2.2
5422
1.7
–––––
4.0
from
08.95
6940
1.7
– Operation Mode (V3H0) "Level manual" extended.
New parameters for the "Level manual" mode:
Manual Level, Table Editing (V3H6),
Row No. Table (V3H7),
Input Level (V3H8,)
Set Volume (V3H9).
– "Pressure before Bias", "Pressure after Bias" and
"Security Locking" parameters moved from the
Linearisation menu to the Service menu.
– Error and warning list extended.
– "Set Simulation Current" (V7H1) parameter,
Simulations limits from 3.8 to 22 mA.
– "MIN Alarm" option for "Set Output Safety" (V0H8)
parameter no longer possible.
5.0
from
12.98
6950
1.8
– "MIN Alarm" option for "Set Output Safety"(V0H8) parameter
available again.
– "Set Simulation Current" (V7H1):
Simulation limits again from 3.6 to 22 mA
Note!
Cerabar S
Software History
Hauser
5