Installation and commissioning
Biometra TRIO
20
Version 03_19
Make sure that the main supply voltage is in accordance with the label above the
power connection (see chapter "Operation Voltage" p. 20).
Connect the Biometra TRIO Thermal Cycler to a grounded socket.
Prior to connecting the unit to the power source, ensure that the voltage selector at
the back side of the instrument is set to the required voltage.
WARNING
Danger of electric shock! Unplug the power cable before you open the Biometra TRIO
Thermal Cycler!
The display contrast can be adjusted to local light conditions.
4.4
Operation Voltage
WARNING
Danger of electric shock! Prior to connecting the Biometra TRIO Thermal Cycler to the
mains, make sure that the setting of the voltage selector is in accordance with your
mains voltage.
The Biometra TRIO Thermal Cycler can operate at 100, 115 or 230 Volt. The operation
Voltage is shown on the Voltage selector, which is located at the instrument bottom.
To change operation voltage, switch off the instrument and disconnect the mains plug.
Use a coin or another round shape item to turn the adjustment slot of the Voltage
selector to the new Voltage.