HOOSE BLOCK START PROGRAM
By activating the checkboxes select the sample blocks for program start.
Fig. 17 Multiblock selection query
Note:
Programs offered for quick start can only be edited or deleted by users with the
right W
RITE
/D
ELETE OWN PROGRAMS
activated (see section "Editing user settings" p. 96).
E
NABLE A USER TO EDIT OR
DELETE OWN PROGRAMS
Press the corresponding button to activate the programming mode or select a program
from the list of templates:
E
DIT
N
EW FROM TEMPLATE
6.7
Incubation mode
Note:
Programs can only be started by users with the right S
TART
/
STOP PROGRAMS
activated (see section "Editing user settings" p. 96).
E
NABLE A USER TO START
OR STOP PROGRAMS
The incubation mode allows samples to be incubated at a constant temperature for an
indefinite period of time. Press the [I
NCUBATE
] button and in the next screen set the
values for the block temperature and hold time using the keypad. Set whether the
heated lid should be on or off and whether it should preheat or not using the
corresponding buttons (see Fig. 18).