Page 5-2
card. There are no user serviceable assemblies in the 6634C Vibration Amplifier. If a card is determined to be faulty,
return it to ENDEVCO following the policy stated below.
A.
FACTORY SERVICE
ENDEVCO maintains factory service for repair of all ENDEVCO instruments. To obtain repair or replacement of a
defective instrument, return the instrument to the factory at the following address:
ENDEVCO
30700 Rancho Viejo Road
San Juan Capistrano, CA 92675
949-493-8181
To return an instrument, please:
1-
Ensure proper packaging, preferably in the original shipping cartons.
2-
Prepay shipping and insurance charges.
B.
WARRANTY RETURNS
ENDEVCO warrants each new electronic instrument to be free from defects in material and workmanship. Please
refer to the appropriate Warranty Policies for terms and warranty period. Instruments returned under warranty
will be repaired or replaced at no charge if the failure is due to defective material or workmanship and the unit is
returned within the time period and conditions of ENDEVCO's warranty policy.
C.
OUT-OF-WARRANTY-RETURNS
Enclose a purchase order not to exceed $300 authorizing repair/calibration along with a full explanation of the
failure or symptoms observed to help in fixing the problem. A copy of a Report of Failure Checklist is attached
immediately following this section to assist in describing the failure that occurred. The instrument will be
repaired/calibrated, returned and invoiced for the actual repair charges. If the instrument cannot be repaired for
$300 or less, ENDEVCO will submit a written quotation for the cost of the repair and/or replacement of the unit.
3.
REPLACING EPROMS
Firmware upgrades are implemented by installing a new EPROM on the main circuit board.
To replace the 27256 EPROM (U38 on the Main Board), remove power by unplugging power cord from rear of unit, and
remove cover by removing 4 head-head screws in top surface of unit.
WARNING: HANDLE UNCOVERED UNIT ONLY AFTER GROUNDING YOUR BODY WITH AN ESD DEVICE, SUCH
AS A GROUNDED WRIST STRAP.
Locate the 27256 (U38) EPROM as shown in Figure 5-1. Using a DIP IC removal tool, pull U38 from its socket and store
it in anti-static wrapping. Check the pin alignment on the replacement IC. Straighten pins if necessary. Orient the
EPROM as shown in Figure 5-1. Install the IC by carefully lining up the pins with the socket holes and pressing the IC into
the socket.
Replace the cover and tighten the four screws through the top surface of the cover into the unit.