ENDEVCO 6634C
INSTRUCTION MANUAL
IM6634C
Page 4-4
If new input sensitivity is successfully setup, no response will be provided by the 6634C. If a problem is
encountered, and unit is in the Listener/Talker mode, the following message will be provided:
E1XXXX "BEL" Setup entered is not allowed; unit will proceed with previous settings
After the new sensitivity is successfully implemented, the unit returns to Normal Mode. Different sensitivities can
be selected for each input type. The new sensitivity settings are stored as the defaults for each input type in non-
volatile memory.
The allowable sensitivity ranges are as follows:
PE (single-ended or differential)
1.500 to 150.0 pC/g
Velocity Coil, or DRCC VEL out
15.00 to 1500 mV/ips
RCC Input
1.500 to 150.0 mV/g
DRCC ACC out
(SENS x DRCC GAIN)/10
DRCC VEL out
15.00 to 1500 mV/ips
To determine the current input sensitivity setting, send the following command to the 6634C:
S?XXXX ; Request current input sensitivity
Unit responds (if unit is in Listener/Talker mode) with:
Snnnnn "ACK" where nnnnn is sensitivity in floating point decimal format.
The sensitivity is indicated in the appropriate units for the selected input type, as follows:
pC/g
Single-ended or Differential PE
mV/ips
velocity coil
mV/g
RCC input (optional)
3.
OUTPUT SETTINGS
The Model 6634C can be remotely set to provide outputs in a variety of engineering units.
At power-up, the outputs are set to the default settings which were previously stored in non-volatile memory.
A.
MODE
To remotely set the Mode (acceleration, velocity, or displacement), send one of the following commands to the
6634C:
M0XXXX ;
Set unit to acceleration output mode
M1XXXX ;
Set unit to velocity output mode
M2XXXX ;
Set unit to displacement output mode
If new output mode is successfully setup, no response will be provided by the 6634C. If a problem is encountered,
and the unit is in the Listener/Talker mode, the error message E1XXXX "BEL", setup is not allowed will be
provided by the 6634C.