Figure 4-4: Ceramic Diffuser Element Replacement
A. Threaded probe flange
B. M5-0.8 x 5 mm locking set screw (use 2.5 mm hex key)
C. Spanner wrench
D. Ceramic diffusion element
E. M5-1 x 6 mm setscrew (use 3 mm hex key)
F. Vee deflector
G. Cement fillet
H. Cement port
I. Flame arrestor hub
4. If equipped with the optional ceramic diffusion assembly, remove and discard the
setscrews and remove the vee deflector (
) to turn the hub free from the retainer. Inspect
the diffusion element. If damaged, replace the element.
5. Loosen the four socket head cap screws from the cell and flange assembly.
6. Remove the assembly and the corrugated cell.
Reference Manual
Maintenance and service
00809-0100-4880
October 2019
Rosemount Oxymitter DR
61
Содержание Rosemount Oxymitter DR
Страница 1: ...Reference Manual 00809 0100 4880 Rev AB October 2019 Rosemount Oxymitter DR In Situ Oxygen Probe ...
Страница 4: ...4 ...
Страница 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Страница 66: ...Maintenance and service Reference Manual October 2019 00809 0100 4880 66 Emerson com Rosemount ...
Страница 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Страница 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Страница 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...
Страница 94: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4880 94 Emerson com Rosemount ...
Страница 95: ...Reference Manual Declaration of Conformity 00809 0100 4880 October 2019 Rosemount Oxymitter DR 95 ...
Страница 96: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4880 96 Emerson com Rosemount ...
Страница 97: ...E China RoHS table Reference Manual China RoHS table 00809 0100 4880 October 2019 Rosemount Oxymitter DR 97 ...