Figure 2-20: Rosemount Oxymitter DR probe wired to the ZA8C or AV8C converter
Heater temperature set to 1380 °F (750 °C).
The greater mass of the Rosemount Oxymittter DR probe requires a longer time to heat
up. Upon startup, the Yokogawa electronics may indicate an error, because the probe has
not reached temperature setpoint in the normal time. Remove power from the Yokogawa
electronics or probe module to clear the error and restore power. You may have to repeat
this procedure a couple of times before probe operating temperature is reached.
A. ZA8C or AV8C terminal
B. Cell
C. Thermocouple
D. Cold junction
E. Ground shield
F. ZA8C or AV8C
G. Probe heater
H. ZA8C or AV8C converter
I. Heater and signal cable
J. Rosemount Oxymitter DR probe
K. Existing mounting
Reference Manual
Install
00809-0100-4880
October 2019
Rosemount Oxymitter DR
49
Содержание Rosemount Oxymitter DR
Страница 1: ...Reference Manual 00809 0100 4880 Rev AB October 2019 Rosemount Oxymitter DR In Situ Oxygen Probe ...
Страница 4: ...4 ...
Страница 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Страница 66: ...Maintenance and service Reference Manual October 2019 00809 0100 4880 66 Emerson com Rosemount ...
Страница 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Страница 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Страница 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...
Страница 94: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4880 94 Emerson com Rosemount ...
Страница 95: ...Reference Manual Declaration of Conformity 00809 0100 4880 October 2019 Rosemount Oxymitter DR 95 ...
Страница 96: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4880 96 Emerson com Rosemount ...
Страница 97: ...E China RoHS table Reference Manual China RoHS table 00809 0100 4880 October 2019 Rosemount Oxymitter DR 97 ...