Relative humidity
5 to 95% (non-condensing)
Signals
Analog output/HART: 4-20 mA isolated from power supply, 950
ohms maximum load
Logic I/O: Two terminal logic contact configurable as either an
alarm output or as a bi-directional calibration handshake signal
to Rosemount IMPS 4000 or SPS 4001B, self-powered (+5V) in
series with 340 ohms conduit ports -
¾
-in. -14 NPT (for analog
output and logic I/O signal lines)
Power requirements
Normal operation: 175 W
Maximum (warmup): 500 W
Ambient operating
temperature (junction box)
200 °F (93 °C)
160 °F (71 °C) maximum for YEW replacement
Rosemount has satisfied all obligations coming from the
European legislation to harmonize the product requirements in
Europe.
1.11.2
Electrical specifications
Rosemount
™
Oxymitter power requirements and consumption
Probe heater electrical power
100 to 240 V ±10%, 50/60 Hz
Power consumption of probe heater
776 VA maximum during warm-up
Electrical power of Rosemount
Oxymitter or optional Xi electronics
120 to 240 V ±10%, 50/60 Hz
Power consumption of Rosemount
Xi
10 watts maximum
Transmitter electrical power
12 to 42 Vdc (loop-powered from the control
room or from the Rosemount Xi box)
Rosemount Xi power requirements and consumption
Electrical power
100 to 240 V ±10%, 50 -60 Hz
Power consumption
Traditional architecture: 776 VA maximum
Digital architecture: 12 VA maximum
Rosemount SPS 4001B power requirements
100 to 240 Vac, 50/60 Hz, 15 VA
Rosemount IMPS 4000 power requirements
85 to 264 Vac, 50/60 Hz, 50 VA
Description and specifications
Reference Manual
October 2019
00809-0100-4340
34
Emerson.com/Rosemount
Содержание Rosemount Oxymitter 4000
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Страница 84: ...Configure transmitter with LOI Reference Manual October 2019 00809 0100 4340 84 Emerson com Rosemount ...
Страница 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Страница 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Страница 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...
Страница 188: ...Safety data sheet for ceramic fiber products Reference Manual October 2019 00809 0100 4340 188 Emerson com Rosemount ...
Страница 192: ...Rosemount Oxymitter product certifications Reference Manual October 2019 00809 0100 4340 192 Emerson com Rosemount ...
Страница 194: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 194 Emerson com Rosemount ...
Страница 195: ...Reference Manual Declaration of Conformity 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 195 ...
Страница 196: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 196 Emerson com Rosemount ...
Страница 197: ...G China RoHS table Reference Manual China RoHS table 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 197 ...