6.8
LOI installation
The LOI connects to the top of the electronic assembly in the electronics housing.
There are four matching connectors on the back of the LOI module that allow the LOI to be
oriented as desired.
Figure 6-7: LOI Module Rear View - LOI Connector Receptacles
A. Connector receptacles
6.9
Test points
Test points 1 through 6 allow you to monitor the following with a multimeter: the heater
thermocouple, the O
2
cell millivolt value, and the process O
2
.
•
TP1 and TP2 monitor the oxygen cell millivolt output, which equates to the percentage
of oxygen present.
•
TP3 and TP4 monitor the heater thermocouple.
•
TP5 and TP6 monitor the process gas or the calibration gas procedures.
Reference Manual
Start up and operate transmitter with LOI
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
101
Содержание Rosemount Oxymitter 4000
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Страница 84: ...Configure transmitter with LOI Reference Manual October 2019 00809 0100 4340 84 Emerson com Rosemount ...
Страница 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Страница 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Страница 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...
Страница 188: ...Safety data sheet for ceramic fiber products Reference Manual October 2019 00809 0100 4340 188 Emerson com Rosemount ...
Страница 192: ...Rosemount Oxymitter product certifications Reference Manual October 2019 00809 0100 4340 192 Emerson com Rosemount ...
Страница 194: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 194 Emerson com Rosemount ...
Страница 195: ...Reference Manual Declaration of Conformity 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 195 ...
Страница 196: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 196 Emerson com Rosemount ...
Страница 197: ...G China RoHS table Reference Manual China RoHS table 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 197 ...