
8
Troubleshoot
8.1
Overview
While the transmitter electronics provides a significant number of diagnostic alarms to
assist in troubleshooting potential problems, it is important to place these alarms in
perspective with respect to the instrument's operating principles.
When the zirconium oxide sensing cell is heated to its setpoint (1357 °F [736 °C]), the cell
will generate a voltage that represents the difference between the process O
2
percentage
and the reference O
2
percentage inside the probe (20.95 percent O
2
ambient air).
Test points (
) are provided to read the raw millivolt value generated by the
thermocouple that controls both the cell temperature and the raw cell signal.
Figure 8-1: O
2
Sensor mV Readings vs. Percent O
2
at 1357 °F (736 °C) (Reference Air,
20.95 Percent O
2
)
Table 8-1: Logic I/O Configuration (as Set at HART/AMS or LOI)
O
2
%
EMF (mV)
100
-34
20
1.0
15
7.25
10
16.1
Reference Manual
Troubleshoot
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
115
Содержание Rosemount Oxymitter 4000
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Страница 84: ...Configure transmitter with LOI Reference Manual October 2019 00809 0100 4340 84 Emerson com Rosemount ...
Страница 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Страница 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Страница 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...
Страница 188: ...Safety data sheet for ceramic fiber products Reference Manual October 2019 00809 0100 4340 188 Emerson com Rosemount ...
Страница 192: ...Rosemount Oxymitter product certifications Reference Manual October 2019 00809 0100 4340 192 Emerson com Rosemount ...
Страница 194: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 194 Emerson com Rosemount ...
Страница 195: ...Reference Manual Declaration of Conformity 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 195 ...
Страница 196: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 196 Emerson com Rosemount ...
Страница 197: ...G China RoHS table Reference Manual China RoHS table 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 197 ...