Figure 8-10: HEATER LED on Membrane Keypad
Recommended actions
1. Remove power.
2. Remove the electronic assembly per .
3. Using a multimeter, measure the resistance across the terminals of heater
connector, J8.
The measurement should be approximately 72 ohms. If the heater is open, see
Fault 5, LOI
When Fault 5 is detected, the LOI displays the
O2 Heater Open
message.
Reference Manual
Troubleshoot
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
127
Содержание Rosemount Oxymitter 4000
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Страница 84: ...Configure transmitter with LOI Reference Manual October 2019 00809 0100 4340 84 Emerson com Rosemount ...
Страница 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Страница 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Страница 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...
Страница 188: ...Safety data sheet for ceramic fiber products Reference Manual October 2019 00809 0100 4340 188 Emerson com Rosemount ...
Страница 192: ...Rosemount Oxymitter product certifications Reference Manual October 2019 00809 0100 4340 192 Emerson com Rosemount ...
Страница 194: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 194 Emerson com Rosemount ...
Страница 195: ...Reference Manual Declaration of Conformity 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 195 ...
Страница 196: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 196 Emerson com Rosemount ...
Страница 197: ...G China RoHS table Reference Manual China RoHS table 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 197 ...