
Cal constants - Results
of the calibration
Current Calibration
If the calibration passed, the transmitter updates these values.
Log these levels into the calibration record (
). if the process has high levels of particulate, the
transmitter removes the response back to the process after
cal gas.
Previous Calibration
Values from the prior good calibration.
Failed Calibration
The transmitter does not load bad calibration values into the
electronics.
Cal constants status
Calibration Step
The current step in an active calibration procedure.
Calibration Time
Time until the next scheduled calibration.
Next O2 Cal
Time until the next O
2
calibration, if different than the next
scheduled calibration.
9.3
Repair
Each of the following procedures details how to remove and replace a specific component
of the transmitter.
WARNING
The transmitter is hot and may cause severe burns.
Remove the transmitter from the stack for all service activities. Allow the transmitter to
cool and take it to a clean work area.
WARNING
Electrical shock
Disconnect and lock out power before working on any electrical components. There is
voltage up to 115 Vac.
9.3.1
Remove probe
Refer to
if you are using integral electronics or
remote electronics. The numbers indicated in the steps below correspond to the numbers
of the items in those figures.
Procedure
1. Turn off power to the system.
2. Shut off the calibration gases at the cylinders and instrument air.
Reference Manual
Maintenance and service
00809-0100-4340
October 2019
Rosemount Oxymitter 4000
157
Содержание Rosemount Oxymitter 4000
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual October 2019 00809 0100 4340 viii Emerson com Rosemount ...
Страница 84: ...Configure transmitter with LOI Reference Manual October 2019 00809 0100 4340 84 Emerson com Rosemount ...
Страница 110: ...HART AMS Reference Manual October 2019 00809 0100 4340 110 Emerson com Rosemount ...
Страница 114: ...HART AMS Reference Manual October 2019 00809 0100 4340 114 Emerson com Rosemount ...
Страница 184: ...Optional accessories Reference Manual October 2019 00809 0100 4340 184 Emerson com Rosemount ...
Страница 188: ...Safety data sheet for ceramic fiber products Reference Manual October 2019 00809 0100 4340 188 Emerson com Rosemount ...
Страница 192: ...Rosemount Oxymitter product certifications Reference Manual October 2019 00809 0100 4340 192 Emerson com Rosemount ...
Страница 194: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 194 Emerson com Rosemount ...
Страница 195: ...Reference Manual Declaration of Conformity 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 195 ...
Страница 196: ...Declaration of Conformity Reference Manual October 2019 00809 0100 4340 196 Emerson com Rosemount ...
Страница 197: ...G China RoHS table Reference Manual China RoHS table 00809 0100 4340 October 2019 Rosemount Oxymitter 4000 197 ...