1
Description and specifications
1.1
Specifications
Rosemount
™
1056 Transmitter
For Rosemount 1056 Transmitter specifications, see the Rosemount 1056 Transmitter
Reference Manual on
Manual: Rosemount 1056 Dual-Input
Transmitter
.
Table 1-1: General Specifications
Characteristic
Specification
Sample requirements
•
Pressure: 3 to 65 psig (122 to 549 kPa abs). A
check valve in the inlet prevents the sensor
flow cells from going dry if sample flow is
lost. The check valve opens at 3 psig (122
kPa abs). If the check valve is removed,
minimum pressure is 1 psig (108 kPa abs).
•
Temperature: 32 to 122 °F (0 to 50 °C)
•
Minimum flow: 3 gal/hr (11 L/hr)
•
Maximum flow: 80 gal/hr (303 L/hr); high
flow causes the overflow tube to back up.
Sample conductivity
>50 µS/cm at 77 °F (25 °C)
Process connection
¼
-in. OD tubing compression fitting (can be
removed and replaced with barbed fitting for
soft tubing)
Drain connection
¾
-in. barbed fitting. Sample must drain to open
atmosphere.
Wetted parts
Overflow sampler and flow cell: acrylic,
polycarbonate, Kynar
®
, nylon, and silicone
Chlorine sensor: Noryl
®
, Viton
®
, wood, silicone,
polyethersulfone, polyester, and platinum
pH sensor (Rosemount
™
3900VP): Stainless
steel, glass, Teflon
®
, polyphenylene sulfide,
EPDM, and silicone
Response time to step change in chlorine
concentration
< 80 sec to 95% of final reading for inlet sample
flow of 3 gph (11 L/hr)
Weight/shipping weight (rounded up to nearest
1 lb. or 0.5 kg)
Rosemount FCL-01: 10 lb./13 lb. (4.5 kg/6.0 kg)
Rosemount FCL-02: 11 lb./14 lb. (5.0 kg/6.5 kg)
Table 1-2: Sensor Specifications
Characteristic
Specification
Free chlorine range
0 to 10 ppm as Cl
2
. For higher ranges, consult
the factory.
Reference Manual
Description and specifications
00809-0100-3412
May 2019
Rosemount FCL 1056
7
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...