If the zero current is moderately larger than expected, an error message appears.
9. To force the transmitter to accept the zero current, choose Yes. To repeat the
calibration, choose No.
If the zero current is much larger than expected, the
Sensor zero failed
screen
appears.
The transmitter will not update the zero current.
10. To return to the main display, press
MENU
and then
EXIT
.
6.3.3
Calibrate the sensor
Procedure
1. Place the chlorine sensor in the chlorine flow cell.
2. If continuous (live) pH correction is being used, calibrate the pH sensor (Section)
and place it in the pH flow cell. If manual pH correction is being used, measure the
pH of the sample and enter the value.
See
.
3. Adjust the chlorine concentration until it is near the upper end of the operating
range. Wait until the transmitter reading is stable before starting calibration.
4. Press
MENU
.
The main
Menu
screen appears. The cursor is on Calibrate.
Reference Manual
Calibrate
00809-0100-3412
May 2019
Rosemount FCL 1056
67
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...