Filter
The transmitter applies a software filter to the raw voltage value
coming from the pH sensor. The filter reduces noise, but increases the
response time. See
Reference
impedance
Usually, the impedance of the reference electrode in a pH sensor is
low. However, a few pH sensors have high reference impedance, and
the transmitter must be told that the reference impedance is high. The
pH sensor used in the Rosemount FCL-02 has low reference
impedance.
5.5.4
Configure measurement
Complete the following steps to configure the transmitter to measure free chlorine.
Procedure
1. Press MENU.
The main
Menu
screen appears.
2. Move the cursor to Program and press
ENTER
.
3. Choose Measurement.
The screen below appears only if you have a Rosemount
™
FCL-02.
Choose Sensor 1 (chlorine) or Sensor 2 (pH).
The screen summarizes the present configuration for sensor 1 (chlorine). If you
have a Rosemount FCL-02, the items are Measure, Units, Filter, Free Cl Correct, and
Resolution. If you have a Rosemount FCL-01, the items are Measure, Units, Filter,
Manual pH, and Resolution. Only four items are shown at a time.To view the
remaining items, scroll to the bottom of the screen and continue scrolling.
4. To make a change, move the cursor to the desired line and press
ENTER
.
A screen appears in which the present setting can be edited.
5. To store the setting, press
ENTER
.
a) .For Measurement, choose Free Chlorine. Do not choose pH Independ. Free
Cl.
b) Leave Filter at the default value (5 sec) unless readings are noisy.
Reference Manual
Programming the transmitter
00809-0100-3412
May 2019
Rosemount FCL 1056
55
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...