Figure 5-3: Operation of the Interval Timer
A. On time duration (0 - 999 sec)
B. On (relay activated)
C. Timer interval (0 - 999.9 hr)
D. Recovery (0 - 999 sec)
E. Hold
The numbers in parentheses are the allowed values for
each timer parameter.
Synchronize
timer
If two or more relays are being used as interval timers, choosing
synchronize timers will cause each timer to start one minute later than
the preceding timer.
5.4.3
Configure alarms and assign setpoints
The Rosemount
™
FCL has an optional alarm relay board. This section describes how to
configure and assign setpoints to the alarm relays, simulate alarm action, and synchronize
interval timers.
Important
Configure the alarms first.
1. Configuring an alarm means
a. Assigning a sensor and measurement to an alarm. An alarm relay can also be
used as a timer.
b. Selecting high or low logic.
c. Choosing the deadband.
d. Setting the interval timer parameters.
2. Simulating an alarm means making the transmitter energize or de-energize an
alarm relay.
Procedure
1. Press
MENU
.
Reference Manual
Programming the transmitter
00809-0100-3412
May 2019
Rosemount FCL 1056
49
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...