b) Use the navigation keys to change
10.00
to
8.5
ppm. Use
Left
and
Right
to
move from digit to digit. Use
Up
and
Down
to increase or decrease the
numeral.
c) To move the decimal point, press
Left
or
Right
until the decimal point is
highlighted. Press
Up
to move the decimal point to the right. Press
Down
to
move to the left.
d) Press
ENTER
to store the setting.
The display returns to the summary screen shown below. Note that the 20 mA
setting for output 1 has changed to
8.50 ppm
.
6. To return to the main menu, press
MENU
. To return to the main display, press
MENU
and then
EXIT
.
4.4
Security
4.4.1
How the security code works
Security codes prevent accidental or unwanted changes to program settings or
calibrations. There are three levels of security.
1. A user can view the default display and diagnostic screens only.
2. A user has access to the calibration and hold menus only.
3. A user has access to all menus.
Display and operation
Reference Manual
May 2019
00809-0100-3412
34
Emerson.com/Rosemount
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...