UNDER NO CIRCUMSTANCES
must the analyzer be started with the protector cap
fitted as this will prevent the analyzer from venting during the purge process.
WARNING
FLAMMABLE/EXPLOSIVE GASES
The purge/pressurization cycle expels any flammable or explosive gases from the
analyzer's electrical compartment. The analyzer must be purged for a minimum of 2
minutes 30 seconds at a minimum flow rate of 280 L/m at 1.5 ± 0.5 bar to prevent
hazards to personnel during the startup process.
Only after the purge/pressurization cycle has been completed can electrical power
be applied to the rest of the analyzer.
10. Turn on the compressed air supply to the analyzer.
11. Set the main circuit breaker to
On
. The air pressurization will begin its automatic
start-up and begin to purge the air inside the analyzer.
Figure 7-2: Purge Control Panel
A. K1 LED
B. P/SV LED
C. PT100 LED
12. Observe the PT100 LED on the purge control panel (
Start-up procedure
Reference Manual
October 2019
00809-0100-4512
78
Rosemount CT5100(Ex)
Содержание Rosemount CT5100
Страница 1: ...Reference Manual 00809 0100 4512 Rev AA October 2019 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 16: ...16 ...
Страница 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Страница 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Страница 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Страница 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Страница 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Страница 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Страница 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...
Страница 172: ...ATEX IECEX assessment report summary Reference Manual October 2019 00809 0100 4512 172 Rosemount CT5100 Ex ...