4. Disconnect the wiring harness (H).
5. Remove and retain the four screws (B) nuts (G), plain washers (D) and spring
washers (F).
6. Remove and discard the unserviceable detector module (C).
7. Inspect the new detector module for damage.
8. Fit the new detector module in position on the detector bracket (E). Secure the
detector module with the four screws, nuts and washers retained in
G).
9. Connect the wiring harness (H) to the detector module (C).
10. Connect the trigger OUT harness (A) to the detector module (C).
11. Close the door of the electrical compartment and restart the analyzer.
12.15
Replacing the O
2
detector
The O
2
detector is located on the underside of the optical compartment base plate. First
determine which detector type is fitted: 1 or 2. Replace the O
2
detector identified using
the following steps.
12.15.1
Replacing O
2
detector (type 1)
Procedure
1. Shut down the analyzer. Refer to
2. Open the door of the electrical compartment.
3. Disconnect the Bayonet Neill–Concelman (BNC) cable and the Power cable from the
O
2
detector (D).
Reference Manual
Repair and replacement
00809-0100-4512
October 2019
Reference Manual
147
Содержание Rosemount CT5100
Страница 1: ...Reference Manual 00809 0100 4512 Rev AA October 2019 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 16: ...16 ...
Страница 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Страница 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Страница 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Страница 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Страница 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Страница 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Страница 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...
Страница 172: ...ATEX IECEX assessment report summary Reference Manual October 2019 00809 0100 4512 172 Rosemount CT5100 Ex ...