2. Open the door of the electrical compartment.
3. Tag or otherwise identify all wiring harnesses before disconnecting.
4. Disconnect the 12 V power wiring harness (E) from the peripheral board (A).
NOTICE
The other end of the 12 V power wiring harness (H) is connected to connector J11
on the motherboard.
A ribbon cable (E) that connects the TEC board to the motherboard runs across the
top of the peripheral board and is secured with a cable tie (C) to the connector of
the ribbon cable (B) that plugs into the peripheral board.
5. Cut the cable tie (C) that secures the TEC board ribbon cable (E) to the connector of
the peripheral board ribbon cable (B). Carefully move the TEC board ribbon cable
clear of the peripheral board.
6. Disconnect the peripheral board ribbon cable (B) from the peripheral board.
7. Disconnect the relay wiring harness (I) from the peripheral board.
8. Disconnect the three connectors (J, K, and L) of the gas/cell temperature
monitoring harness from the peripheral board.
9. Remove and retain the four screws (H) and associated spring washers (G) and flat
washers (F).
10. Remove and discard the peripheral board.
11. Place the peripheral board (A) in position on the backplate.
12. Secure the peripheral board by fitting the four screws (H) and the associated spring
washers (G) and flat washers (F) retained during the removal procedure.
13. Torque tighten the screws to 0.6 Nm.
14. Connect the three connectors (J, K, and L) of the gas/cell temperature monitoring
harness to, respectively, connectors J9, J24, and J37 on the peripheral printed circuit
card (PCC).
15. Connect the relay wiring harness (I) to the peripheral board.
16. Connect the peripheral board ribbon cable (B) to the peripheral board.
17. Secure the TEC board ribbon cable (D) to the connector of the peripheral board
ribbon cable (B) with a cable tie (C).
18. Connect the 12 V power wiring harness (E) to the peripheral board (A).
19. Close the door of the electrical compartment and restart the analyzer.
Repair and replacement
Reference Manual
October 2019
00809-0100-4512
136
Rosemount CT5100(Ex)
Содержание Rosemount CT5100
Страница 1: ...Reference Manual 00809 0100 4512 Rev AA October 2019 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Страница 16: ...16 ...
Страница 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Страница 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Страница 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Страница 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Страница 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Страница 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Страница 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...
Страница 172: ...ATEX IECEX assessment report summary Reference Manual October 2019 00809 0100 4512 172 Rosemount CT5100 Ex ...