Wafer sensors installation components and assembly requirements
Figure 3-7:
A.
Ground ring (optional)
B.
Customer supplied gaskets
C.
Spacer installation (horizontal meters)
D.
Spacer installation (vertical meters)
E.
O-ring
F.
Installation studs, nuts, and washers (optional)
G.
Wafer alignment spacer
H.
Flow
Gaskets
The sensor requires a gasket at each process connection. The gasket material selected
must be compatible with the process fluid and operating conditions. Gaskets are required
on each side of a grounding ring. See
Note
Metallic or spiral-wound gaskets should not be used as they will damage the liner face of the sensor.
Alignment spacers
On 1.5 inch through 8 inch (40 through 200 mm) line sizes, Rosemount requires installing
the alignment spacers to ensure proper centering of the wafer sensor between the process
flanges. To order an Alignment Spacer Kit (quantity 3 spacers) use p/n 08711-3211-xxxx
where xxxx equals the dash number shown in
Rosemount alignment spacers
Table 3-5:
Dash-no. (-xxxx)
Line size
Flange rating
(in)
(mm)
0A15
1.5
40
JIS 10K-20K
0A20
2
50
JIS 10K-20K
0A30
3
80
JIS 10K
Sensor Installation
Reference manual
17
Содержание Rosemount 8732EM
Страница 2: ......
Страница 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 87: ...Field Communicator Dashboard Menu Tree HART v5 4 part 2 Figure 7 8 Operation Reference manual 81 ...
Страница 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 138: ...Advanced Diagnostics Configuration 132 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Страница 221: ...Wiring Diagram 475 Field Communicator with 8732EM External Analog Power Figure C 6 Wiring Diagrams Reference manual 215 ...
Страница 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 253: ...Implementing a Universal Transmitter Reference manual 247 ...