Depending on the method used to view the results, they will be displayed in either a menu
structure, as a method, or in the report format. When using the HART
®
Field
Communicator, each individual parameter can be viewed as a menu item. When using the
LOI, the parameters are viewed as a method using the left arrow key to cycle through the
results. In AMS, the calibration verification report is populated with the necessary data
eliminating the need to manually complete the form.
When using AMS there are two possible methods that can be used to print the report:
•
Method one involves using the print functionality within the EDDL screen. This print
functionality essentially prints a screen shot of the report.
•
Method two involves using the print feature within AMS while on the Maintenance
Service Tools screen. This will result in a printout of all of the maintenance
information. Page one of the report contains the meter verification result data.
The results are displayed in the order found in the table below. Each parameter displays a
value used by the diagnostic to evaluate meter health.
Manual Smart Meter Verification Test Parameters
Table 9-3:
Parameter
LOI menu path ( Diagnostics > Varia-
bles > MV Results > Manual Results )
1
Test Condition
Test Condition
2
Test Criteria
Test Criteria
3
8714i Test Result
MV Results
4
Simulated Velocity
Sim Velocity
5
Actual Velocity
ActualVelocity
6
Velocity Deviation
Flow Sim Dev
7
Xmtr Cal Test Result
Xmtr Cal Verify
8
Sensor Cal Deviation
Sensor Cal Dev
9
Sensor Cal Test Result
Sensor Cal
10
Coil Circuit Test Result
Coil Circuit
11
Electrode Circuit Test Result
Electrode Ckt
Continuous SMART Meter Verification Test Parameters
Table 9-4:
Parameter
LOI menu path ( Diagnostics > Varia-
bles > MV Results > Continual Res )
1
Continuous Limit
Test Criteria
2
Simulated Velocity
Sim Velocity
3
Actual Velocity
ActualVelocity
4
Velocity Deviation
Flow Sim Dev
5
Coil Signature
Coil Inductnce
6
Sensor Cal Deviation
Sensor Cal Dev
7
Coil Resistance
Coil Resist
Advanced Diagnostics Configuration
126
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Содержание Rosemount 8732EM
Страница 2: ......
Страница 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 87: ...Field Communicator Dashboard Menu Tree HART v5 4 part 2 Figure 7 8 Operation Reference manual 81 ...
Страница 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 138: ...Advanced Diagnostics Configuration 132 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Страница 221: ...Wiring Diagram 475 Field Communicator with 8732EM External Analog Power Figure C 6 Wiring Diagrams Reference manual 215 ...
Страница 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 253: ...Implementing a Universal Transmitter Reference manual 247 ...