4.
Press when all changes are complete to save the entered values.
5.
Press again to navigate back to the menu tree.
7.2.3
Data entry examples
Parameter values are classified as
table values
or
select values
.
•
Table values are available from a predefined list for parameters such as line size or
flow units.
•
Select values are integers, floating point numbers, or character strings and are
entered one character at a time using the arrow keys for parameters such as PV URV
and calibration number.
Table value example
Setting the sensor size:
Procedure
1.
Press key to access the menu. See
2.
Use , , , and to select line size from the basic setup menu.
3.
Use or to increase/decrease the sensor size.
4.
When you reach the desired sensor size, press .
5.
Set the loop to manual if necessary, and press again.
After a moment, the LOI will display
VALUE STORED SUCCESSFULLY
and then display
the selected value.
Select value example
Changing the upper range limit:
Procedure
1.
Press key to access the menu. See
2.
Use , , , and to select PV URV from the basic setup menu.
3.
Press to position the cursor.
4.
Press or to set the number.
5.
Repeat steps 3and 4 until desired number is displayed, press .
6.
Set the loop to manual if necessary, and press again.
After a moment, the LOI will display
VALUE STORED SUCCESSFULLY
and then display
the selected value.
Operation
Reference manual
71
Содержание Rosemount 8732EM
Страница 2: ......
Страница 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 87: ...Field Communicator Dashboard Menu Tree HART v5 4 part 2 Figure 7 8 Operation Reference manual 81 ...
Страница 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 138: ...Advanced Diagnostics Configuration 132 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Страница 221: ...Wiring Diagram 475 Field Communicator with 8732EM External Analog Power Figure C 6 Wiring Diagrams Reference manual 215 ...
Страница 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 253: ...Implementing a Universal Transmitter Reference manual 247 ...