Set timeline resolution
To set the resolution of the level trend timeline:
Procedure
1. Select
Service Tools
→
Echo Curve
.
2. Select
Options
.
3. In the
Timeline Resolution
list, select the desired length (in hours) of the timeline.
4. Select
Save
.
5. Select
Back
.
7.4.2
Perform an analog loop test
During a loop test, the transmitter outputs a fixed value (4 mA, 20 mA, or user-selected
value).
The loop test command verifies the output of the transmitter, the integrity of the loop,
and the operations of any recorders or similar devices installed in the loop. A loop test can
also be used to determine the need for an analog output calibration.
Procedure
1. Select
Service Tools
→
Simulate
.
2. Under
Analog Out
, select
Loop test
, and then follow the on-screen instructions.
3. Measure the loop current.
Postrequisites
Select
Stop
or
Cancel
to exit loop test (
Abort
on handheld communicator).
Related information
Calibrate analog out
Simulation/test active
7.4.3
Calibrate analog out
Use this function to calibrate the analog output by comparing the actual output current
with the nominal 4 mA and 20 mA currents. Calibration is done at factory and the analog
output does not normally need to be recalibrated.
Procedure
1. Select
Service Tools
→
Maintenance
→
Routine Maintenance
.
2. Select
D/A trim
(
Calibrate Analog Out
in Rosemount Radar Master Plus) and follow
the on-screen instructions.
Service and troubleshooting
Reference Manual
September 2022
00809-0100-4418
90
Emerson.com/Rosemount
Содержание Rosemount 3408
Страница 4: ...4 ...
Страница 14: ...Transmitter overview Reference Manual September 2022 00809 0100 4418 14 Emerson com Rosemount ...
Страница 28: ...Mechanical installation Reference Manual September 2022 00809 0100 4418 28 Emerson com Rosemount ...
Страница 38: ...Electrical installation Reference Manual September 2022 00809 0100 4418 38 Emerson com Rosemount ...
Страница 46: ...Configuration Reference Manual September 2022 00809 0100 4418 46 Emerson com Rosemount ...
Страница 54: ...Operation and maintenance Reference Manual September 2022 00809 0100 4418 54 Emerson com Rosemount ...
Страница 104: ...Service and troubleshooting Reference Manual September 2022 00809 0100 4418 104 Emerson com Rosemount ...