Reference Manual
00809-0100-4716, Rev JA
May 2008
Rosemount 3095 MultiVariable
5-10
The sensing module is a complete assembly and cannot be further
disassembled.
REASSEMBLY
PROCEDURES
Follow these procedures carefully to ensure proper reassembly:
Attaching the Sensor
Module to the
Electronics Housing
1.
Inspect all cover and housing (non-process-wetted) O-rings and
replace if necessary. Lightly lubricate with silicone to ensure a good
seal.
2.
Carefully tuck the cable connector completely inside the internal
shroud. To do this, turn the shroud and cable counterclockwise one
rotation to tighten the cable.
3.
Lower the electronics housing onto the module, and guide the internal
shroud and cable through the housing and into the external shroud.
4.
Fasten the housing to the module by turning clockwise.
NOTE
To prevent damage to the cable connector, watch the cable and shroud as
you attach the housing to the module. Make sure the cable connector does
not slip out of the internal shroud and begin to rotate with the housing.
Reinsert the cable connector into the shroud if it escapes before the housing
is fully fastened.
5.
Inspect the threaded connections.
NOTE
The bottom of the electronic housing must be within
1
/
16
-in. of the sensor
module to maintain explosion-proof requirements.
6.
Tighten the housing rotation set screw.
Attaching the Electronics
Board
1.
Remove the cable connector from its position inside the internal
shroud, and attach the cable to the electronics board.
2.
Align the post-receptacle connectors with the posts inside the
electronics housing.
NOTE
If the post-receptacle connectors have a black rubber sleeve over them, the
sleeve must be removed before installing a new electronics board. Gently
grasp the sleeve between thumb and forefinger and slide it off the connector.
Discard sleeve.
3.
Insert the electronics board into the housing and tighten the captive
mounting screws.
4.
Replace the electronics housing cover. Metal to metal contact is
preferred.
Содержание Rosemount 3095
Страница 2: ......
Страница 8: ......
Страница 130: ...Reference Manual 00809 0100 4716 Rev JA May 2008 Rosemount 3095 MultiVariable 5 24 ...