a) Select the
XD_SCALE_UNITS
from the dropdown menu.
b) Enter the
XD_SCALE 0%
point.
This may be elevated or suppressed for level applications.
c) Enter the
XD_SCALE 100%
point.
This may be elevated or suppressed for level applications.
d) If
L_TYPE
is
Direct
, you may place the AI block in
AUTO
mode to return the
device to service.
Guided setup does this automatically.
4. If
L_TYPE
is
Indirect
or
Indirect Square Root
, set
OUT_SCALE
to change
engineering units.
a) Select the
OUT_SCALE UNITS
from the dropdown menu.
b) Set the
OUT_SCALE
low value.
This may be elevated or suppressed for level applications.
c) Set the
OUT_SCALE high
value.
This may be elevated or suppressed for level applications.
d) If
L_TYPE
is
Indirect
, you may place the AI Block in
AUTO
mode to return the
device to service.
Guided Setup does this automatically.
5. If
L_TYPE
is
Indirect Square Root
, a
LOW FLOW CUTOFF
function is available.
a) Enable
LOW FLOW CUTOFF
.
b) Set the
LOW_CUT VALUE
in
XD_SCALE UNITS
.
c) You may place the AI Block in
AUTO
mode to return the device to service.
Guided Setup does this automatically.
6. Change damping.
•
To use guided setup:
—
Navigate to
Configure > Guided Setup
, and select Change Damping.
Note
Guided setup automatically goes through each step in the proper order.
—
Enter the desired damping value in seconds. The permitted range of values is
0.4 to 60 seconds.
•
To use manual setup:
—
Navigate to
Configure > Manual Setup > Process Variable
, and select
Change Damping
.
—
Enter the desired damping value in seconds. The permitted range of values is
0.4 to 60 seconds.
7. Configure optional LCD display (if installed).
Reference Manual
Configuration
00809-0100-4774
July 2020
Reference Manual
37
Содержание Rosemount 3051 CF Series
Страница 4: ...4 ...
Страница 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Страница 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Страница 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Страница 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Страница 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Страница 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...