b. Hold the flange adapters and adapter O-rings in place while installing (in
the desired of the four possible process connection spacing
configurations) using four 2.88-in. bolts to mount securely to the
coplanar flange. For gage pressure configurations, use two 2.88-inch
bolts and two 1.75-in. bolts.
•
Manifold: Contact the manifold manufacturer for the appropriate bolts and
procedures.
3. Tighten the bolts to the initial torque value using a crossed pattern.
See
for appropriate torque values.
4. Using the same cross pattern, tighten bolts to final torque values seen in
.
Table 6-1: Bolt Installation Torque Values
Bolt material
Initial torque value
Final torque value
CS-ASTM-A445 Standard
300 in.-lb (34 N-m)
650 in.-lb (73 N-m)
316 SST—Option L4
150 in.-lb (17 N-m)
300 in.-lb (34 N-m)
ASTM-A-19 B7M—Option L5
300 in.-lb (34 N-m)
650 in.-lb (73 N-m)
ASTM-A-193 Class 2, Grade B8M—Option L8
150 in.-lb (17 N-m)
300 in.-lb (34 N-m)
Note
If you replaced the PTFE sensor module O-rings, re-torque the flange bolts after
installation to compensate for cold flow of the O-ring material.
Note
For Range 1 transmitters, after replacing O-rings and re-installing the process flange,
expose the transmitter to a temperature of 185 °F (85 °C) for two hours. Then re-tighten
the flange bolts in a cross pattern and again expose the transmitter to a temperature of
185 °F (85 °C) for two hours before calibration.
for complete warning information.
6.4.4
Install drain/vent valve
Complete the following steps to reinstall the drain/vent valve in the Rosemount 3051
Transmitter.
Procedure
1. Apply sealing tape to the threads on the seat. Starting at the base of the valve with
the threaded end pointing toward the installer, apply five clockwise turns of sealing
tape.
2. Tighten the drain/vent valve to 250 in.-lb. (28.25 N-m).
3. Take care to place the opening on the valve so that process fluid will drain toward
the ground and away from human contact when the valve is opened.
Troubleshooting
Reference Manual
July 2020
00809-0100-4774
108
Rosemount 3051
Содержание Rosemount 3051 CF Series
Страница 4: ...4 ...
Страница 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Страница 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Страница 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Страница 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Страница 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Страница 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...