Field Communicator Menu Trees and Fast Keys
January 2018
Reference Manual
00809-0400-4101, Rev BA
Field Communicator Menu Trees and Fast Keys
75
B.2
Fast Keys
A (
) indicates the basic configuration parameters. At minimum these parameters should be verified
as a part of configuration and startup.
A (7) indicates availability only in HART revision 7 mode.
Table B-1. Device Revision 9 and 10 (HART7), DD Revision 1 Fast Key Sequence
Function
Fast Key sequence
HART 7
HART 5
Alarm and Saturation Levels
2, 2, 2, 5
2, 2, 2, 5
Damping
2, 2, 1, 1, 5
2, 2, 1, 1, 5
Primary Variable
2, 2, 5, 1, 1
2, 2, 5, 1, 1
Range Values
2, 2, 2, 1
2, 2, 2, 1
Tag
2, 2, 7, 1, 1
2, 2, 7, 1, 1
Transfer Function
2, 2, 1, 1, 6
2, 2, 1, 1, 6
Pressure Units
2, 2, 1, 1, 4
2, 2, 1, 1, 4
Date
2, 2, 7, 1, 5
2, 2, 7, 1, 4
Descriptor
2, 2, 7, 1, 6
2, 2, 7, 1, 5
Digital to Analog Trim (4 - 20 mA Output)
3, 4, 2, 1
3, 4, 2, 1
Digital Zero Trim
3, 4, 1, 3
3, 4, 1, 3
Display Configuration
2, 2, 4
2, 2, 4
LOI Password Protection
2, 2, 6, 5
2, 2, 6, 4
Loop Test
3, 5, 1
3, 5, 1
Lower Sensor Trim
3, 4, 1, 2
3, 4, 1, 2
Message
2, 2, 7, 1, 7
2, 2, 7, 1, 6
Pressure Trend
3, 3, 1
3, 3, 1
Rerange with Keypad
2, 2, 2, 1
2, 2, 2, 1
Scaled D/A Trim (4 - 20 mA / 1-5 V) Output)
3, 4, 2, 2
3, 4, 2, 2
Scaled Variable
2, 2, 3
2, 2, 3
Sensor Temperature Trend
3, 3, 3
3, 3, 3
Switch HART Revision
2, 2, 5, 2, 4
2, 2, 5, 2, 3
Upper Sensor Trim
3, 4, 1, 1
3, 4, 1, 1
7
Long Tag
2, 2, 7, 1, 2
N/A
7
Locate Device
3, 4, 5
N/A
7
Simulate Digital Signal
3, 5
N/A
Содержание Rosemount 2051G
Страница 2: ......
Страница 10: ...2 Reference Manual 00809 0400 4101 Rev BA Introduction January 2018 Introduction ...
Страница 34: ...26 Reference Manual 00809 0300 4410 Rev BA Configuration January 2018 Configuration ...
Страница 64: ...56 Reference Manual 00809 0400 4101 Rev BA Operation and Maintenance January 2018 Operation and Maintenance ...
Страница 72: ...64 Reference Manual 00809 0400 4101 Rev BA Troubleshooting January 2018 Troubleshooting ...
Страница 89: ......