25
Reference Manual
00809-0300-4410, Rev BA
Configuration
January 2018
Configuration
2.14.1
Changing a transmitter address
To activate multidrop communication, the transmitter poll address must be assigned a number from 1
to 15 for HART Revision 5, and 1–63 for HART Revision 7. Each transmitter in a multidropped loop must
have a unique poll address.
Field Communicator
AMS Device Manager
1. Right click the device and select
Configure
.
2. In HART Revision 5 mode:
a. Select
Manual Setup
, select the
HART
tab.
b. In the Communication Settings box enter polling address in the
Polling Address
box, select
Send
.
3. In HART Revision 7 mode:
a. Select
Manual Setup
, select the
HART
tab and select the
Change Polling Address
button.
4. Carefully read the warning and click
Yes
if it is safe to apply the changes.
2.14.2
Communicating with a multidropped transmitter
To communicate with a multidrop transmitter, the Field Communicator or AMS Device Manager has to
be set up for Polling.
Field Communicator
1. Select
Utility
and
Configure HART Application
.
2. Select
Polling Addresses.
3. Enter
0–63.
AMS Device Manager
1. Select the
HART modem icon
and select
Scan All Devices
.
From the
HOME
screen, enter the Fast Key sequence
HART Revision 5
HART Revision 7
Device Dashboard Fast Keys
2, 2, 5, 2, 1
2, 2, 5, 2, 2
Содержание Rosemount 2051G
Страница 2: ......
Страница 10: ...2 Reference Manual 00809 0400 4101 Rev BA Introduction January 2018 Introduction ...
Страница 34: ...26 Reference Manual 00809 0300 4410 Rev BA Configuration January 2018 Configuration ...
Страница 64: ...56 Reference Manual 00809 0400 4101 Rev BA Operation and Maintenance January 2018 Operation and Maintenance ...
Страница 72: ...64 Reference Manual 00809 0400 4101 Rev BA Troubleshooting January 2018 Troubleshooting ...
Страница 89: ......