55
Reference Manual
00809-0400-4101, Rev BA
Operation and Maintenance
January 2018
Operation and Maintenance
5.7.3
AMS Device Manager
1. Select
Manual Setup
and select
HART.
2. Select
Change HART Revision,
then follow the on screen prompts.
Note
AMS Device Manager versions 10.5 or greater are compatible with HART Revision 7.
5.7.4
LOI
Navigate to
HART REV
within the extended menu and select either
HART REV 5
or
HART REV 7
. Use
below to change HART Revision.
Figure 5-8. Change HART Revision with LOI
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
HART REV
BACK TO MENU
EXIT MENU
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU