Section 5: Operation and Maintenance
April 2013
The Sensor Calibration and the Analog Output Calibration combine to match the transmitter’s
measurement system to the plant standard.
Calibrate the sensor
Sensor Trim (
Zero Trim (
Calibrate the 4-20 mA output
4-20 mA/ 1-5V Output Trim (
)
4-20 mA/ 1-5V Output Trim using other scale (
5.4.1
Determining necessary Sensor Trims
Bench calibrations allow for calibrating the instrument for its desired range of operation.
Straight forward connections to pressure source allow for a full calibration at the planned
operating points. Exercising the Transmitter over the desired pressure range allows for
verification of the analog output.
Trim the pressure signal on page 71
discusses how the trim
operations change the calibration. It is possible to degrade the performance of the transmitter
if a trim is done improperly or with inaccurate equipment. The transmitter can be set back to
factory settings using the Recall Factory Trim command in
Recall Factory Trim—sensor trim on
.
For transmitters that are field installed, the manifolds discussed in
allow the differential transmitter to be zeroed using the zero trim
function. Both 3-valve and 5-valve manifolds are discussed. This field calibration will eliminate
any pressure offsets caused by mounting effects (head effect of the oil fill) and static pressure
effects of the process.
Determine the necessary trims with the following steps.
1.
Apply Pressure
2.
Check digital pressure, if the digital pressure does not match the applied pressure,
perform a digital trim. See
Perform a Sensor Trim on page 72
3.
Check reported analog output against the live analog output. If they do not match,
perform an analog output trim. See
Performing Digital-to-Analog Trim (4-20mA/ 1-5 V
Trimming with configuration buttons
Local configuration buttons are external buttons located underneath the top tag of the
transmitter. There are two possible sets of local configuration buttons that can be ordered and
used to perform trim operations: Digital Zero Trim and Local Operator Interface. To access the
buttons, loosen screw and rotate top tag until buttons are visible.
Local Operator Interface (M4):
Can perform both digital Sensor Trim and the 4-20mA
Output Trim (analog output trim). Follow the same procedures listed in trimming with
Field Communicator or AMS Device Manager listed below.
Digital Zero Trim (DZ):
Used for performing a sensor zero trim. See
calibration frequency on page 68
for trim instructions.
Содержание Rosemount 2051
Страница 2: ......
Страница 4: ......
Страница 102: ...92 Reference Manual 00809 0100 4107 Rev BA Section 6 Troubleshooting April 2013 Troubleshooting ...
Страница 213: ...203 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 214: ...204 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 227: ......