7.3.2
Level is stuck in measuring range
Figure 7-2: Symptom
A
B
C
D
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Disturbing object in the tank.
Recommended actions
•
Analyze the Echo Peaks and check General Threshold.
•
Remove the disturbing object.
•
Put an inclined metal plate on top of the disturbing object.
•
Move the transmitter to another position.
Related information
Analyze the echo peaks
Adjust the general threshold