3.2.6
Nozzle requirements
To allow the microwaves to propagate undisturbed, the nozzle dimensions should be kept
within the specified limits as given in
. The inside of the nozzle must be smooth
(i.e. avoid bad welding, rust, or deposit).
Figure 3-5: Mounting in Nozzles
D
H
Table 3-3: Nozzle Requirements
Nozzle diameter (D)
Maximum nozzle height (H)
1 in. (25 mm)
3.9 in. (100 mm)
1.5 in. (40 mm)
5.9 in. (150 mm)
2 in. (50 mm)
7.9 in. (200 mm)
3 in. (80 mm)
11.8 in. (300 mm)
4 in. (100 mm)
15.8 in. (400 mm)
6 in. (150 mm)
23.6 in. (600 mm)
3.2.7
Open air installations
The antenna extension must be fitted to the transmitter to comply with open air
requirements. The antenna extension is included in the bracket mount option.
Figure 3-6: Antenna Extension
Related information
Reference Manual
Mechanical installation
00809-0200-4480
May 2021
Reference Manual
17
Содержание Rosemount 1408A
Страница 6: ...Introduction Reference Manual May 2021 00809 0200 4480 6 Rosemount 1408A Level and Flow Transmitter ...
Страница 12: ...Transmitter overview Reference Manual May 2021 00809 0200 4480 12 Rosemount 1408A Level and Flow Transmitter ...
Страница 89: ...Reference Manual 00809 0200 4480 May 2021 Reference Manual 89 ...