Oxymitter 5000
9-4
Instruction Manual
IM-106-350, Rev 2.2
July 2008
Figure 9-2. Inside Right Cover
2. Verify the calibration gas parameters are correct per "Calibration with
3. If performing a manual calibration with the CALIBRATION
RECOMMENDED LED off and the CAL LED off, start at step a.
4. If performing a manual calibration with the CALIBRATION
RECOMMENDED LED on and the CAL LED on, start at step b.
a. Push the CAL key. The CALIBRATION RECOMMENDED LED will
come on and the CAL LED will be on solid. If a multimeter is
attached across TP5 and TP6, the reading will display the
percentage of oxygen seen by the cell.
b. Push the CAL key. The CALIBRATION RECOMMENDED LED will
turn off and the CAL LED will flash continuously. The flashing LED
indicates that the Oxymitter 5000 is ready to accept the first
calibration gas.
c. Apply the first calibration gas. (Electronics will abort the calibration if
step d is not done within 30 minutes).
d. Push the CAL key; the CAL LED will be on solid. A timer is activated
to allow the calibration gas adequate time to flow (default time of five
minutes). When the timer times out, the Oxymitter 5000 has taken
the readings using the first calibration gas and the CAL LED will flash
continuously. The flashing indicates the Oxymitter 5000 is ready to
take readings using the second calibration gas.
e. Remove the first calibration gas and apply the second calibration
gas. (Electronics will abort the calibration if step f is not done within
30 minutes).
SW2 DIP SWITCH
OFF
ON
NOT USED
NOT USED
LED
HEATER
O2 CELL
CALIBRATION
HEATER T/C
STATUS
OPEN
SHORTED
REVERSED
HIGH HIGH
TEMP
HIGH CASE
TEMP
LOW TEMP
HIGH TEMP
OPEN
BAD
EPROM
CORRUPT
INVALID SLOPE
INVALID
CONSTANT
FLASHES
1
2
3
4
1
2
OPEN
3
4
5
1
3
4
1
2
PUSH CAL
CAL LED ON
APPLY TG1
PUSH CAL
CAL LED ON SOLID
REMOVE TG1 & APPLY TG2
PUSH CAL
CAL LED ON SOLID
WAIT FOR FLASH
REMOVE TG2
PUSH CAL
1
2
3
4
5
6
7
ALARMS
MANUAL
CALIBRATION
CAL LED ON FOR
PURGE TIME
WAIT FOR FLASH
CAL LED OFF
PUSH CAL
CAL LED FLASH
2 FLASH-VALID CAL
3 FLASH-INVALID CAL
8
PLACE CONTROL LOOP
IN MANUAL
IF CAL LED ON
GO TO STEP 2
*
*
3
LAST CAL
FAILED
A/D COMM
ERROR
38730038
NOT USED
NOT USED
NOT USED
NOT USED
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...