Oxymitter 5000
9-2
Instruction Manual
IM-106-350, Rev 2.2
July 2008
Figure 9-1. Membrane Keypad
calibration gas flowmeter to 5 scfh and calibrate the Oxymitter 5000. To
change the diffusion element refer to "Ceramic Diffusion Element
Replacement".
Three types of calibration methods are available: automatic, semi-automatic,
and manual.
NOTE
A calibration can be aborted any time during the process. Press the CAL key
(Figure 9-1) on the Oxymitter 5000 keypad three times within three seconds,
or abort via the LOI, FOUNDATION fieldbus, or an IMPS 4000. An aborted
calibration will retain the values of the previous good calibration.
Automatic Calibration
Automatic calibrations require no operator action. However, the calibration
gases must be permanently piped to the Oxymitter 5000, an SPS 4001B or
IMPS 4000 must be installed to sequence the gases, and the logic I/O must
be set to mode 8 via the FOUNDATION fieldbus computer terminal using the
IO_PIN_MODE parameter so the sequencer and Oxymitter 5000 can commu-
nicate.
Depending on your system setup, an automatic calibration can be initiated by
the following methods:
1. The Oxymitter 5000's CALIBRATION RECOMMENDED alarm signals
that a calibration is required.
2. Use the LOI to enter a "time since last cal" (CAL INTRVL) parameter
that will initiate an automatic calibration at a scheduled time interval (in
hours). To configure the CAL INTRVL parameter, refer to "Oxymitter
5000 Setup at the LOI" in Section 6: Startup and Operation of Oxymitter
5000 with LOI for the LOI.
38730014
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATER T/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
INC
INC
DEC
DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Страница 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...