Instruction Manual
IM-106-880, Rev 1.0
January 2007
6-9
OCX 8800
5. From the
SELECT ACTION
screen, select
START/NEXT CALSTEP
to
continue calibration, select
ABORT CAL
to abort calibration or
EXIT
CAL
to exit calibration. Select one from the list and press
ENTER
.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the
AppO2Low
step, switch on O
2
Low Gas. Verify the O
2
concentration measured matches the
O2 LOW
GAS
parameter in
Setup
. Press
OK
when ready.
7. Select
Start/Next Cal Step
to start applying the O
2
Low Gas. The time
to apply the test gas is specified by the
Gas Time
.
8. The Calibration Status should automatically change to
FIowO2Low
and
then
ReadO2Low
for a period of time. During this period, if an attempt
is made to go to the next calibration step by pressing
OK
and selecting
Start/Next Cal Step
, you will be prompted with
Operator step
command is not accepted at this time
. The
Next Cal Step
command
is not accepted at this time.
9. When ready, Calibration Status will stop at
AppO2Hi
. Switch off the O
2
Low Gas and switch on the O
2
High Gas. Verify the O
2
concentration
measured matches the
O2 HIGH GAS
parameter in
Setup
. Press
OK
when ready.
10. Select
Start/Next Cal Step
to apply the O
2
High Gas. The time to apply
the test gas is specified by the
Gas Time
.
11. The Calibration Status should automatically change to
FlowO2Hi
, then
ReadO2Hi
, and then
ReadCOeLo
for a period of time. During this
period, if an attempt is made to go the next calibration step by pressing
OK
and selecting
Start/Next Cal Step
, you will be prompted with
Operator step command is not accepted at this time
. The
Next Cal
Step
command is not accepted at this time.
12. When ready, Calibration Status will stop at
AppCOeHi
. Switch off the O
2
High Gas and switch on the COe Gas. Verify the COe concentration
measured matches the
COe TEST GAS
parameter in the
Setup
. Press
OK
when ready.
13. Select
Start/Next Cal Step
to start applying the COe Gas. The time to
apply the test gas is specified by the
Gas Time
.
14. The Calibration Status should automatically change to
FlowCOeHi
and
then
ReadCOeHi
for a period of time. During this period, if an attempt is
made to go the next calibration step by pressing
OK
and selecting
Start/Next Cal Step
, you will be prompted with
Operator step
command is not accepted at this time
. The
Next Cal Step
command
is not accepted at this time.
15. When ready, Calibration Status will stop at
STOP GAS
. Switch off the
COe gas. Press
OK
when ready. Select
Start/Next Cal Step
to start
purging gas. The time to purge gas is specified by
Purge Time
.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 25: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 2 5 OCX 8800 Figure 2 2 Adapter Plate Installation ...
Страница 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Страница 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Страница 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Страница 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Страница 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...