Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
6-4
OPERATOR - INITIATED
AUTOCALIBRATION
An operator can initiate an automatic calibration at any time provided that the
unit is equipped with calibration solenoids.
Autocalibration using the LOI
To initiate a calibration using the LOI, perform the following steps on the LOI
menu tree. Refer to Section 4, Using the LOI, for the LOI menu tree.
1. From the
CALIBRATION
menu, use the right-pointing arrow to select
the
Cal Control
menu.
2. Select
Start Cal-O2
,
Start Cal COe
, or
Start Cal Both
to start the
calibration. Select
Cal Verify
to access the calibration window.
3. At the prompt, use the right-pointing arrow to initiate automatic
calibration.
Autocalibration using HART
To initiate an automatic calibration using HART/AMS, perform the following
steps on the HART menu tree. Refer to Section 5, Using HART
Communications, for the HART menu tree.
1. Select
DIAG/SERVICE
from
DEVICE SETUP
menu.
2. Select
CALIBRATE
from the
DIAG/SERVICE
menu.
3. Select
PERFORM CAL
from the
CALIBRATE
menu.
4. Select
CAL METHODS
from the
PERFORM CAL
menu.
5. From the
CAL METHODS
menu, select the type of calibration desired:
O2 Calibration
,
COe Calibration
, or
O2 and COe Calibration
.
MANUAL CALIBRATION
If a unit is not equipped with calibration solenoids, a calibration must be
performed by an operator following prompts from the unit. Refer to the
following paragraphs for manual calibration.
Manual Calibration using the LOI
Use the following procedure to perform a manual calibration with the LOI. If
necessary, refer to the menu tree in Section 4, Using the LOI. Once the
manual calibration procedure is initiated at the LOI, a series of prompts will
appear giving instructions to the operator.
1. Use the right-pointing key to select the
CALIBRATION
first column
submenu.
2. From the
CALIBRATION
submenu use the right-pointing key to select
the
Cal Control
second column submenu.
3. From the
Cal Control
submenu use the right-pointing key to select the
third column
Start Cal O2
option.
4. Remain at the
Start Cal O2
option or use the down-pointing key to
select the
Start Cal COe
option or
Start Cal Both
option. (The following
sequence applies when
Start Cal Both
is selected.)
5. Use the right-pointing key to start the calibration. Turn on the low O
2
test
gas, when prompted by the
Flow Low Gas
message.
6. Press the right-pointing key when the low O
2
test gas is applied. The
calibration data changes as the calibration proceeds.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 25: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 2 5 OCX 8800 Figure 2 2 Adapter Plate Installation ...
Страница 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Страница 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Страница 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Страница 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Страница 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...