OCX 8800
D/A TRIM PROCEDURES
O
2
D/A trim procedure using HART/AMS
Use the following procedure to perform the O
2
D/A trim procedure using the
HART communicator or AMS. Refer to the HART menu tree, Figure 5-3.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the
DIAG/SERVICE
menu, select
D/A TRIM
. The HART
communicator displays
O2 D/A Trim
.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select
ABORT
.)
3. The HART communicator displays
WARNING: Loop should be
removed from automatic control
. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press
OK
.
4. Remove the electronics housing cover.
5. Refer to Figure 2-4. Connect a digital multimeter to read the milliamp
output from the O
2
D/A coverter circuit. Connect the positive lead to the
AOUT1+ terminal and connect the negative lead to the AOUT1-
terminal. Then, press
OK
at the HART communicator.
6. The HART communicator displays
Setting Fld dev output to 4 mA
.
Press
OK
. Read the O
2
millamp output at the digital multimeter. Enter
the reading at the HART communicator and press
ENTER
. (Select
ABORT
to exit without changes).
7. The HART communicator displays
Setting Fld dev output to 20 mA
.
Press
OK
. Read the O
2
millamp output at the digital multimeter. Enter
the reading at the HART communicator and press
ENTER
. (Select
ABORT
to exit without changes).
8. The HART communicator displays
Setting Fld dev output to 4 mA
.
Press
OK
.
9. The HART communicator displays
Fld dev output 4.00 mA
equal to
reference meter?
Using the up or down arrow, select
1 Yes
or
2 No
and Press
ENTER
. If
No
, the process repeats from step 6.
10. The HART communicator displays
Setting Fld dev output to 20 mA
.
Press
OK
.
11. The HART communicator displays
Fld dev output 20.00 mA
equal to
reference meter?
Using the up or down arrow, select
1 Yes
or
2 No
and Press
ENTER
. If
No
, the process repeats from step 7.
12. The HART communicator displays
NOTE: Loop may be returned to
automatic control
.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 25: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 2 5 OCX 8800 Figure 2 2 Adapter Plate Installation ...
Страница 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Страница 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Страница 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Страница 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Страница 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...