Instruction Manual
IM-106-880, Rev 1.0
January 2007
6-3
OCX 8800
4. Select
O2 CalIntrvl
(O
2
calibration interval) and enter the desired time
in hours between automatic O
2
calibrations. Select
COE Callintrvl
and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O
2
and COe, enter 0 for both
CalIntrvl
parameters.
5. If desired, the
O2NxtCalTm
and the
COeNxtCalTm
(next calibration
time) parameters can be changed to synchronize a calibration at a
specific day or time.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
6. From the
O2 CALIB PARAMS
screen, select
CalIntrvl
.
7. At the prompt, input a time interval (in hours) at which an automatic O
2
calibration will occur and press
ENTER
.
8. From the
DETAILED SETUP
screen, select
COE CALIB PARAMS
.
9. From the
COE CALIB PARAMS
menu, select
CalIntrvl
.
10. At the prompt, input a time interval (in hours) at which an automatic COe
calibration will occur and press
ENTER
.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O
2
and COe are NOT calibrated simultaneously.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 25: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 2 5 OCX 8800 Figure 2 2 Adapter Plate Installation ...
Страница 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Страница 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Страница 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Страница 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Страница 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...