![Emerson Machinery Health Expert CSI 2600 Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/emerson/machinery-health-expert-csi-2600/machinery-health-expert-csi-2600_user-manual_103529089.webp)
•
uses a special calibration utility program (DHM).
GS recalibration should be performed:
•
annually.
•
whenever a 6510 Signal Input module is replaced.
•
whenever a 6560 Processor module is replaced.
•
if the calibration table has a status of "Unknown".
Product Support personnel can guide a plant engineer or technician through GS calibration
over the phone.
B.1.4
Digital Signal Processor (DSP) calibration
Digital Signal Processor (DSP) calibration:
•
uses a CPU board's TSG output signal.
•
does not require that any wire harnesses be disconnected.
•
is completed in about 30-40 minutes.
•
does not require any special cables or test equipment.
•
uses a special calibration utility program (DHM).
DSP recalibration should be performed:
•
annually.
•
whenever a 6510 Signal Input module is replaced.
•
whenever a 6560 Processor module is replaced.
•
if the calibration table has a status of “Unknown”.
Product Support personnel can guide a plant engineer or technician through DSP
calibration over the phone.
B.1.5
Transient calibration
The Transient unit includes two processing boards—a Main Processor board and a
Transient board. Both boards include separate Digital Signal Processors. The DSP on the
Transient board uses an internal calibration table, in much the same way that the Main
Processor DSP circuit does. However, it is calibrated separately, and not as part of a DSP
calibration for a Main Processor.
Transient calibration:
•
uses a Processor module's TSG output signal.
•
does not require that any wire harnesses be disconnected.
•
is completed in about 30-40 minutes.
•
does not require any special cables or test equipment.
•
uses a special calibration utility program (DHM).
System calibration
MHM-97453, Rev 1
85
Содержание Machinery Health Expert CSI 2600
Страница 1: ...User Guide MHM 97453 Rev 1 April 2014 CSI 2600 Machinery Health Expert User Guide ...
Страница 8: ...CSI 2600 back view Figure 2 2 Introduction to the CSI 2600 4 MHM 97453 Rev 1 ...
Страница 12: ...Introduction to the CSI 2600 8 MHM 97453 Rev 1 ...
Страница 26: ...Getting started 22 MHM 97453 Rev 1 ...
Страница 64: ...Software configuration 60 MHM 97453 Rev 1 ...
Страница 96: ...Data types 92 MHM 97453 Rev 1 ...
Страница 100: ...D 2 Internal wiring diagram for the CSI 2600 ...
Страница 106: ...Troubleshooting 102 MHM 97453 Rev 1 ...