Online database diagram
Figure 3-4:
Collection criteria
•
Analysis Parameter (AP) Set—Defines a particular way to collect spectral data, and
specifies:
-
the number of lines of resolution.
-
any averaging modes and windowing.
-
whether to be order-based.
-
what FMax to use.
-
what parameters are to be collected.
•
Alarm Limits (AL) Set—Each AL Set is associated with a specific AP Set. There may be
multiple AL Sets defined for any given AP Set to accommodate changing monitoring
conditions. The alarm limit definition determines when alarms occur, data is stored,
and output relays are set.
•
Predicate—A collection predicate is an expression that compares the conditions of
vibration levels, input relay states, and/or machine speeds to determine when data
is collected and transient auto-archives are extracted.
Getting started
16
MHM-97453, Rev 1