Figure 3-10: Base Enclosure wire seal installation
A. Base Enclosure cover and socket head screws
2. Position the wire to prevent counterclockwise rotation of the screws when the seal
wire is taut. Feed the security wire beneath the Transmitter Electronics Enclosure
and through the adjacent socket head screw. Twist the wire, removing all slack and
seal.
3. Cut wire ends to remove excess wire.
3.8
Sealing the unit
The unit should be properly sealed with a sealing compound after electrical connections
have been tested according to the customer's Best Practices schedule. Some areas require
a witnessed Acceptance Test for the installed system and require that the meter run for a
predetermined length of time (approximately one to two weeks) before the unit is sealed.
This allows time to verify all electrical connections are correct, that the meter is accurately
measuring flow and that the meter meets the customer’s installation requirements. See
and
.
Installation manual
Electrical Installation
3-9000-771
May 2019
Installation manual
59