Figure 2-7: Incorrect sling attachment
10. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock
loading ever occurs, the hoist ring must be inspected per manufacturer's
recommendations prior to be placed in any further service. If a proper inspection
cannot be performed, discard the hoist ring.
11. NEVER lift with any device, such as hooks, chains, or cables that could create side
pulls that could damage the ring of the hoist ring.
12. NEVER lift more than the ultrasonic meter assembly including electronics and
transducers with the hoist rings. The only exception is that it is safe to lift the meter
with one ASME B16.5 or ASME B16.47 blind flange bolted to each end flange of the
meter. NEVER use the hoist rings on the meter to lift other components such as
meter tubes, piping or fittings attached to the meter. Doing so will exceed the load
rating of the hoist rings.
13. Remove the hoist rings from the meter after lifting is completed and store them in
an appropriate case or container per their manufacturer's recommendation.
14. Apply heavy lubricant or anti-seize to the threads of the plug bolts and reinstall the
plug bolts to keep the tapped holes free of debris and to prevent corrosion.
Mechanical Installation
Installation manual
May 2019
3-9000-771
36
3818 LNG Liquid Ultrasonic Meter
Содержание Daniel 3818
Страница 2: ...2 ...
Страница 10: ...Contents Installation manual May 2019 3 9000 771 x 3818 LNG Liquid Ultrasonic Meter ...
Страница 26: ...Introduction Installation manual May 2019 3 9000 771 26 3818 LNG Liquid Ultrasonic Meter ...
Страница 60: ...Electrical Installation Installation manual May 2019 3 9000 771 60 3818 LNG Liquid Ultrasonic Meter ...
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ...Engineering Drawings Installation manual May 2019 3 9000 771 82 3818 LNG Liquid Ultrasonic Meter ...