EES-101 Series Ultrasonic Flowmeter
Revision: 3.0.1
Page 50 of 65
0x01 - 0xF7
0x86
2(Register length error)
CRC-16/MODBUS
0x01 - 0xF7
0x86
3(Check code error)
CRC-16/MODBUS
0x01 - 0xF7
0x86
4(The function is not
supported at the moment)
CRC-16/MODBUS
1 byte
1 byte
1 byte
2 byte
9.3.2.4
MODBUS Register Address List(default)
The flowmeter MODBUS Register has a read register and a single write register.
Read Register Address List (use 0x03 performance code to read)
Register
address
Register
Data description
Type
No.
registers*
Remark
$0000
40001
Flow/s - low half word
32 bits real
2
$0001
40002
Flow/s - high half word
$0002
40003
Flow/m - low half word
32 bits real
2
$0003
40004
Flow/m- high half word
$0004
40005
Flow/h - low half word
32 bits real
2
$0005
40006
Flow/h - high half word
$0006
40007
Velocity
–
low half word
32 bits real
2
$0007
40008
Velocity
–
high half word
$0008
40009
Positive total
–
low half word
32 bits uint.
2
$0009
40010
Positive total
–
high half word
$000A
40011
Positive total
–
exponent
16 bits int.
1
$000B
40012
Negative total
—
low half word
32 bits int.
2
$000C
40013
Negative total
—
high half word
$000D
40014
Negative total
—
exponent
16 bits int.
1
$000E
40015
Net total
—
low half word
32 bits int.
2
$000F
40016
Net total
-
high half word
$0010
40017
Net total
—
exponent
16 bits int.
1
$0016
40023
Up signal int
–
low half word
32 bits real
2
0
~
99.9
$0017
40024
Up signal int
–
high half word
$0018
40025
Down signal int
–
low half word
32 bits real
2
0
~
99.9
$0019
40026
Down signal int
–
high half word
$001A
40027
Quality
16 bits int.
1
0
~
99