EES-101 Series Ultrasonic Flowmeter
Revision: 3.0.1
Page 12 of 65
Press
key
Press Ent to start Manual Totalizer, then press Ent to
end Manual Totalizer, press Ent to input Standard
Totalizer to get the final K factor. Complete the
calibration with pressing Ent to store.
P
ress
key
Set "Zero Point" and the same as Menu 42.
Press
key
Display System Error Codes.
2.2
Examples
For example, measuring the diameter of 219mm and pipe wall thickness of 6mm, measuring medium is water,
Pipe Material is carbon steel, No Liner, can be operated as follows:
Step1. Pipe outside diameter:
Press the
keys to enter Window
M11, and enter the pipe outside diameter, and
then press the
key to confirm.
Step2. Pipe wall thickness
Press the
key to enter Window
M12, and enter the pipe wall thickness, and
press the
key to confirm.
Step3. Pipe Material
Press the
keys to enter Window
M14, press the
key, press the
or
key to select Pipe Material, and press the
key to confirm.
Step4. Transducer type
(The transmitter is available for various
transducer types.)
Press the
key to enter M23, and
then press
, next press
or
to
select flow transducer type1. CP037, finally
press
to confirm.
Manual Calibrate
Press Ent When Ready
Set Zero [42
Press ENT When Ready
*R------------- [08
System Normal
Menu
1
1
Enter
Pipe Outer Diameter
219.00 mm
Menu
1
2
Enter
Pipe Wall Thickness
6 mm
Menu
1
4
Enter
^
^
Enter
Pipe Material [14
0. Carbon Steel
Menu
2
3
Enter
^
^
Enter
Transducer Type [23
1. CP037